- Distribution System
Ensuring consistent water quality compliance
Our purified water distribution system is customized for chemical engineering, laboratories, food and beverage, hemodialysis water, high-end electronic ultrapure water, sterile medical device cleaning water, pharmaceutical WFI, biological product and vaccine preparation water, and process water in direct contact with sterile preparations. We provide full-process delivery solutions.
- Circulation Standard
The system adopts turbulent circulation design. Through automatic welding, inner wall argon protection, gradient design and regular disinfection, the pipeline system is free of dead legs, leakage and microbial growth, meeting GMP and relevant industry standards.
- System Configuration
Complete system configuration including circulation pumps, filters, valves, instruments and other equipment. Adopts 316L stainless steel or PVDF piping with polished inner surfaces. Welding uses automatic orbital welding with argon protection to ensure stable and reliable water quality.
-
Key Requirements
Circulation flow rate ≥ 1.5 m/s, slope ≥ 0.5%, branch length ≤ 3 times pipe diameter (D). Regular cleaning and disinfection to ensure water quality meets GMP and relevant industry standards.
Note: disinfection methods are optional (pasteurization, ozonation, chemical sanitization, UV irradiation, and combined disinfection).